The impedance presented by the unbalanced VHF-LISN to each terminal of the EUT AC mains port is different (not symmetrical), resulting in an unbalanced load. This leads to a controlled conversion of differential-mode currents to common-mode currents, which depends on the value of the termination impedances, as it would occur with an unbalanced AC mains power network. This device complies with the standard CISPR 16-1-4 / Ed. 5
| Frequency range (nominal): | 30 MHz – 300 MHz |
| Impedance of live terminal with PE terminal bonded to RG | |
| Magnitude | 160 Ω ± 40 % |
| Phase | 0 º ± 40 º |
| Impedance of neutral terminal to RG with PE terminal bonded to RG | |
| Magnitude | 60 Ω ± 40 % |
| Phase | 0 º ± 40 º |
| Impedance of PE terminal to RG | |
| Magnitude | 90 Ω ± 40 % |
| Phase | >0 º ± 40 º |
| Isolation AE – EuT: | ≥ 40 dB |
| Maximum current: | 16 A |
| Maximum voltage: | 1000 V |
| Connector for EuT and AE: | 3×4 mm safety laboratory jacks |
| Ground connection: | 4 mm laboratory jack |
| Standard: | CISPR 16-1-4 / Ed5 |
| Material of the housing: | Aluminum |
| Dimensions: | 171 x 105 x 55 mm |
| Weight: | typ. 760 g |
| Accessories: | Impedance Measurement Adapter (optional) IMA 8803 |