CU 303447 – Artificial Antenna

Add to Quote
CU 303447 - Artificial antenna acc. to ETSI EN 303 447
CU 303447 – Artificial antenna acc. to ETSI EN 303 447
The artificial antenna corresponds to requirements of ETSI EN 303 447. It is equipped with a section of inductance and resistance, which serves as a defined symmetrical termination for the RMI-transmitter. Two built-in current transformers, accessible through the two BNC-connectors ICM and IDM, provide measurement capabilities for both, common mode and differential mode currents. The conversion factor of the current transformers is 1:1, i.e. x dBµA is indicated as x dBµV in the 50 Ω-system. The frequency response of the current transformers is flat with a negligible ripple of less than ± 0.1 dB above 150 kHz.

Frequency Range:150 kHz… 30 MHz
Nominal Impedance (Current Transformer Outputs):50 Ω
Conversion Ratio Current Transformers:1:1 (0 dB)
Nominal Low Frequency Impedance at RMI-Terminals:200 µH ± 20%
Saturation current:>2.0 A
Impedance at the RMI-Terminals:150 Ω ± 40 Ω
Housing Material:Aluminium
Housing Dimensions:105 x 120 x 50 mm
Weight:ca. 440 g
Connector:2 x BNC, 50 Ω female
RMI-Connector:4 mm Safety laboratory connector

Datasheet CU 303447